Class 1 Sound Level Meter NL-53 / Class 2 Sound Level Meter NL-43
with Extended Function Program NX-43EX Instruction Manual Operation Guide

4.1 Inserting the batteries

WARNING

  • Make sure to correctly insert the batteries with the polarities of the batteries “+” and “-” matching the marks. If the polarities are incorrect, the batteries may explode or leak. If you do not intend to use the device, remove the batteries to prevent leakage. If electrolyte leaking from the batteries gets on your skin or clothes, wash it off immediately with water.
  • Be careful not to get your hands pinched or cut when installing the battery compartment cover.

Important

  • Insert four new batteries of the same type. Using different types of batteries or mixing old and new batteries may cause a failure.

1Remove the battery compartment cover.

Remove the battery compartment cover.

2Insert four AA batteries, ensuring that the orientation matches the “+” and “-” marks.

Insert four AA batteries, ensuring that the orientation matches the “+” and “-” marks.

3Reinstall the cover.

Important

  • To maintain the dustproof and waterproof performance, make sure that the battery compartment cover is closed securely.

Continuous operating time with battery

The continuous operating time with batteries is as follows (at 23℃).

Battery type Continuous operating time Conditions
Alkaline battery Approx. 12 h Manual store, Eco setting ON
Approx. 16 h Auto store NX-43EX, Eco setting ON, Leq calculation interval set to 10 minutes (Lp store OFF)
Ni-MH rechargeable battery Approx. 12 h Manual store, Eco setting ON
Approx. 16 h Auto store NX-43EX, Eco setting ON, Leq calculation interval set to 10 minutes (Lp store OFF)

Note